15.00 ELECTRICAL TESTS | |
| |
15.40 Specifications for measurements either for instruments or equipment | |
| |
Measurement of insulation resistance | |
| Measurement of earth connector efficiency Specification 500.PIS |
|
|
||||||||||||
|
|
||||||||||||
| Measurement of insulation resistance (CEI EN 60335) | ||||||||||||
|
The Insulation measurement consists in applying of a programmed
continuous voltage between the phases and the ground of the product
being tested for a set time and measuring of the current which
flows. The equivalent resistance value (MW) is calculated with the formula V/I . The test voltage is generated by a PWM (Pulse Width Modulation) circuit and it is continuously monitored for the whole duration of the test. If the resistance value (MW = V/I) is less than the programmed value, the test is immediately suspended with a negative "Reject" result. If the measured value is greater than the programmed value for the whole duration of the test, the test ends with a positive "Good" result. |
||||||||||||
| Values which can be set for the test | ||||||||||||
| Possibility of test with 3 different voltages (250 - 500 - 1'000 Vdc). | ||||||||||||
| With nominal test Voltage Vdc 250 | ||||||||||||
|
||||||||||||
| With nominal test Voltage Vdc 500 | ||||||||||||
|
||||||||||||
| With nominal test Voltage Vdc 1'000 | ||||||||||||
|
||||||||||||
| Setting and measurement resolution | ||||||||||||
|
||||||||||||
| Test time sec from 0.2 to 99.9 | ||||||||||||
|
Self-testing If the test voltage moves out of the set tolerance, the test ends with the result "Fault" being displayed |
||||||||||||
| Measurement of earth connector efficiency (CEI EN 60335 - Art. 27.5) | ||||||||||||
|
The measurement of the earth connector efficiency consists
in applying a programmed alternating current between the ground
contact present on the product being tested and any metal part
of the same, for the set time and in measuring the voltage drop.
The ohm value is calculated using the formula V/I. This test is carried out with four cables (2 Voltmeter and 2 Ammeter cables) to eliminate any contact resistance. |
||||||||||||
|
The EN 60335 legislation, art. 27.5 prescribes that this value
must not exceed 0.1 W. The applied open-circuit voltage must be 10-12 Vac. The circulating current in the closed circuit should be ³10 - ³25 A |
||||||||||||
| The test can be carried out in two different ways: | ||||||||||||
|
Type R - By ensuring that, for the entire test time, the value
of the equivalent resistance (W =V/I)
remains between the minimum and maximum set values (EN 60335 legislation, art. 27.5 max. value 0.1 W.) Type V - By making sure that the value of the voltage drop normalized with the test current (at 10A or 25A) lies between the minimum and maximum set values. If the values are included in the programmed range, the test ends with a positive result. The test ends with a negative result if the values are out of range. |
||||||||||||
| Table for max. voltage drop in relation to the cable section according to CEI EN60204-1 table 7 | ||||||||||||
|
||||||||||||
| Values which can be set for the test |
| Test R | |||||||
| Resistance | |||||||
|
* maximum |
W |
from |
0.001 |
|
1.000 |
||
|
* minimum |
W |
from |
0.001 |
|
1.000 |
||
| Resolution |
0.001 |
||||||
| Precision | % +/- |
2 |
|||||
| Nominal current | A |
³ 10 |
|
³ 25 |
|||
| Settling time | sec. (minimum) |
0.50 |
|||||
| Test time | sec. |
from |
0.5 |
|
15 |
||
| Tipo V | |||||||
| Voltage drop | |||||||
|
* maximum |
V ca |
from |
0.01 |
|
12.00 |
||
|
* minimum |
V ca |
from |
0.01 |
|
12.00 |
||
| Resolution | V |
0.01 |
|||||
| Precision | % +/- |
2 |
|||||
| Nominal current | A |
³ 10 |
|
³ 25 |
|||
| Settling time | sec. (minimum ) |
0.50 |
|||||
| Test time | sec. |
from |
0.5 |
|
15.00 |
||
| Self-testing |
| If the current value is less than the smallest set value, the warning "Fault" is displayed independently of the type of test (R or V). |